Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Diabetes Technol Ther ; 26(S3): 24-31, 2024 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-38377317

RESUMO

Background: The MiniMed™ 780G system (MM780G) with Guardian™ 4 sensor includes a 100 mg/dL glucose target (GT) and automated insulin corrections up to every 5 min and was recently approved for use in the United States. In the present study, early real-world MM780G performance and the use of recommended system settings (100 mg/dL GT with an active insulin time of 2 h), by individuals with type 1 diabetes, were evaluated. Methods: CareLink™ personal data uploaded between the launch of the MM780G to August 22, 2023 were aggregated and underwent retrospective analysis (based on user consent) and if users had ≥10 days of continuous glucose monitoring (CGM) data. The 24-h day CGM metrics, including mean glucose, percentage of time spent in (%TIR), above (%TAR), and below (%TBR) target range (70-180 mg/dL), in addition to delivered insulin and closed-loop (CL) exits, were compared between an overall group (n = 7499) and individuals who used recommended settings (each, for >95% of the time). An analysis of the same metrics for MiniMed™ 770G system (MM770G) users (n = 3851) who upgraded to the MM780G was also conducted (paired t-test or Wilcoxon signed-rank test, P < 0.05 considered statistically significant). Results: For MM780G users, CGM use, and time in CL were >90% and all MM780G CGM metrics exceeded consensus-recommended goals. With recommended settings (22% of all users), mean %TIR and %TITR (70-140 mg/dL) were 81.4% and 56.4%, respectively. For individuals who upgraded from the MM770G, %TIR and %TITR increased from 73.2% to 78.3% and 45.8% to 52.6%, respectively, while %TAR reduced from 25.1% to 20.2% (P < 0.001, for all three). CL exits/week averaged <1, for all MM780G users. Conclusions: Early real-world MM780G use in the United States demonstrated a high percentage of time in range with low time above and below range. These outcomes are similar to those observed for real-world MM780G use in other countries.


Assuntos
Glicemia , Diabetes Mellitus Tipo 1 , Estados Unidos , Humanos , Automonitorização da Glicemia , Estudos Retrospectivos , Insulina/uso terapêutico , Insulina Regular Humana , Diabetes Mellitus Tipo 1/tratamento farmacológico , Glucose , Hipoglicemiantes/uso terapêutico , Sistemas de Infusão de Insulina
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...